Learn how to run Quality Control (QC) on the IDEXX Catalyst One Chemistry Analyzer. Step-by-step instructions make it easy.Learn more about Catalyst One Work.. 1. For Catalyst One: Tap the Catalyst One icon on the IDEXX VetLab* Station Home screen and then tap Maintenance > SmartQC. OR For Catalyst Dx: Tap the Catalyst Dx icon on the IDEXX VetLab Station Home screen and then tap Quality Control. 2. On the left side of the screen, use the applicable drop-down lists to select when you
Running Quality Control To Run General Quality Control on the Catalyst One analyzer Tap the Catalyst One icon on the IDEXX VetLab Station Home screen. 2. Tap Maintenance and then tap Quality Control. 3. Tap the quality control lot number you are using and then tap Run QC. Page 28: Maintenance Upgrade the software promptly.. Correct order for running quality control on the catalyst one analyzer Step 1: tap the catalyst one analyzer icon at the bottom of the IDEXX VetLab station Home Screen Step 2: tap maintenance Step3: tap quality control Step 4: select a QC lot Step 5: tap run Step 6: load components in the catalyst one drawer, close the drawer , and press start